Remote Commands
R&S
®
FSW
1046
User Manual 1173.9411.02 ─ 43
*RST:
0
Example:
INP:GAIN:STAT ON
INP:GAIN:VAL 15
Switches on 15 dB preamplification.
Manual operation:
See
INPut<ip>:GAIN[:VALue]
<Gain>
This command selects the gain if the preamplifier is activated (
INP:GAIN:STAT ON
,
The command requires the additional preamplifier hardware option.
Suffix:
<ip>
.
1 | 2
For R&S
FSW85 models with two RF input connectors:
1: Input 1 (1
mm [RF Input] connector)
2: Input 2 (1.85
mm [RF2 Input] connector)
For all other models:
irrelevant
Parameters:
<Gain>
15 dB | 30 dB
The availability of gain levels depends on the model of the
R&S
FSW.
R&S
FSW8/13/26: 15 dB and 30 dB
R&S
FSW43 or higher: 30 dB
All other values are rounded to the nearest of these two.
Default unit: DB
Example:
INP:GAIN:STAT ON
INP:GAIN:VAL 30
Switches on 30 dB preamplification.
Manual operation:
See
14.7.3.4
Scaling the Y-Axis
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
................................................................. 1046
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO ONCE
............................................. 1047
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:MODE
............................1047
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:PDIVision
.................................................. 1047
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
................................................. 1048
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
......................................1048
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
<Range>
This command defines the display range of the y-axis (for all traces).
Note that the command works only for a logarithmic scaling. You can select the scaling
with
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
.
Setting Basic Measurement Parameters