Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
353
User Manual 1173.9240.02 ─ 23
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:MARK:FUNC:NDBD:STAT ON
Turns the n dB Down marker on.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:TIME?
This command queries the position of the n dB down markers on the x-axis when mea-
suring in the time domain.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>, <m>
.
irrelevant
Return values:
<TimeX1>
absolute position in time of the n dB marker to the left of the ref-
erence marker in seconds
<TimeX2>
absolute position in time of the n dB marker to the right of the
reference marker in seconds
Example:
INIT:CONT OFF
Switches to single sweep mode
CALC:MARK:FUNC:NDBD ON
Switches on the n dB down function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:NDBD:TIME?
Outputs the time values of the temporary markers.
Usage:
Query only
Manual operation:
See
Phase Noise Measurement Marker
The following commands control the phase noise measurement marker function.
CALCulate<n>:MARKer<m>:FUNCtion:PNOise[:STATe]
................................................... 353
CALCulate<n>:MARKer<m>:FUNCtion:PNOise:RESult?
................................................... 354
CALCulate<n>:MARKer<m>:FUNCtion:PNOise[:STATe]
<State>
This command turns the phase noise measurement at the marker position on and off in
the Analog Demodulation application.
Analyzing Results