Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
211
User Manual 1173.9240.02 ─ 23
Parameters:
<PeakVoltage>
0.25 V | 0.5 V | 1 V | 2 V
Peak voltage level at the connector.
For probes, the possible full scale values are adapted according
to the probe's attenuation and maximum allowed power.
*RST:
1V
Example:
INP:IQ:FULL 0.5V
Manual operation:
See
"Full Scale Level Mode / Value"
INPut:IQ:TYPE
<DataType>
This command defines the format of the input signal.
Parameters:
<DataType>
IQ | I | Q
IQ
The input signal is filtered and resampled to the sample rate of
the application.
Two input channels are required for each input signal, one for
the in-phase component, and one for the quadrature compo-
nent.
I
The in-phase component of the input signal is filtered and
resampled to the sample rate of the application. If the center fre-
quency is not 0, the in-phase component of the input signal is
down-converted first (Low IF I).
Q
The quadrature component of the input signal is filtered and
resampled to the sample rate of the application. If the center fre-
quency is not 0, the quadrature component of the input signal is
down-converted first (Low IF Q).
*RST:
IQ
Example:
INP:IQ:TYPE Q
Manual operation:
See
CALibration:AIQ:HATiming[:STATe]
<State>
Activates a mode with enhanced timing accuracy between analog baseband, RF and
external trigger signals.
For more information see the R&S
FSW I/Q Analyzer and I/Q Input User Manual.
Configuring the Measurement