Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
929
User Manual 1178.8520.02 ─ 01
Manual operation:
See
[SENSe:]PROBe<pb>:SETup:PMOFfset
<PMOffset>
Sets the P-mode offset. The setting is only available if a modular probe in P-mode is
connected to the R&S
FSV/A. The maximum voltage difference between the positive
and negative input terminals is 16
V.
If the probe is disconnected, the P-mode offset of the probe is reset to 0.0
V.
Note that if the offset for DM-mode or CM-mode is changed, the offsets for the P-mode
and N-mode are adapted accordingly, and vice versa.
For details see
"MultiMode Function and Offset Compensation for Modular RF Probes"
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Parameters:
<PMOffset>
The voltage offset between the positive input terminal and
ground.
Default unit: V
Manual operation:
See
"Common Mode Offset / Diff. Mode Offset / P Offset / N Off-
[SENSe:]PROBe<pb>:SETup:STATe?
Queries if the probe at the specified connector is active (detected) or not active (not
detected).
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Return values:
<State>
DETected | NDETected
Example:
//Query connector state
PROB3:SET:STAT?
Usage:
Query only
[SENSe:]PROBe<pb>:SETup:TYPE?
Queries the type of the probe.
Setting Basic Measurement Parameters