General Instrument Setup
R&S
®
FSVA3000/ R&S
®
FSV3000
582
User Manual 1178.8520.02 ─ 01
Rule State
Enables/disables processing of the defined rule for the currently active channel only.
Note that the application does not perform a validity check on the rules. For example, if
you change or delete the required command file for a rule, the rule becomes invalid.
However, the application still attempts to process the rule as far as possible. An easy
way to detect such a situation is using the log information, see
ging Rules with the EBA-Journal"
Rule
Name of the rule
Events
Specified trigger events for the rule
Actions
Specified actions to be performed by the rule.
Triggered
Number of trigger events for the rule that occurred since the last instrument preset or
restart.
New Rule
Opens the "Event Based Actions: Modify Rule" dialog box to define a new rule as
described in
Chapter 12.3.2, "Defining Rules"
Edit Rule
Opens the "Event Based Actions: Modify Rule" dialog box to edit the selected rule as
described in
Chapter 12.3.2, "Defining Rules"
Copy Rule
Creates a copy of the selected rule and opens the "Event Based Actions: Modify Rule"
dialog box to edit it, as described in
Chapter 12.3.2, "Defining Rules"
Delete Rule
Deletes the selected event-based action rule.
Show Journal
Shows the logging information for the rules defined in the EBA Rule Manager. This
information is useful to debug rules when expected actions are not performed. For
details, see
Chapter 12.3.3, "Debugging Rules with the EBA-Journal"
12.3.2
Defining Rules
Access
: "Event-based actions" > "New Rule" / "Edit Rule"
Rules are defined globally for all measurement channels. However, you must activate
each rule for each application individually, and only the currently active channel pro-
cesses the rules.
Event-based Actions