Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
357
User Manual 1178.8520.02 ─ 01
"User Defined" Sends a trigger when you select the "Send Trigger" button.
In this case, further parameters are available for the output signal.
Remote command:
Level ← Output Type ← Trigger 1/2
Defines whether a high (1) or low (0) constant signal is sent to the trigger output con-
nector.
The trigger pulse level is always opposite to the constant signal level defined here. For
example, for "Level = High", a constant high signal is output to the connector until you
select the
function. Then, a low pulse is provided.
Remote command:
Pulse Length ← Output Type ← Trigger 1/2
Defines the duration of the pulse (pulse width) sent as a trigger to the output connector.
Remote command:
OUTPut:TRIGger<tp>:PULSe:LENGth
Send Trigger ← Output Type ← Trigger 1/2
Sends a user-defined trigger to the output connector immediately.
Note that the trigger pulse level is always opposite to the constant signal level defined
by the output
setting. For example, for "Level" = "High", a constant high signal is
output to the connector until you select the "Send Trigger" function. Then, a low pulse
is sent.
Which pulse level will be sent is indicated by a graphic on the button.
Remote command:
OUTPut:TRIGger<tp>:PULSe:IMMediate
8.2.9
How to Output a Trigger Signal
Using the variable Trigger 1/2 connector of the R&S
FSV/A, the internal trigger signal
can be output for use by other connected devices. For details on the connectors see
the R&S
FSV/A "Getting Started" manual.
To output a trigger to a connected device
1. Select [Trigger] > "Trigger Config".
2. Switch to the "Trigger In/Out" tab of the "Trigger and Gate".
Data Input and Output