Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
272
User Manual 1178.8536.02 ─ 01
FORMat:DEXPort:HEADer
<State>
If enabled, additional instrument and measurement settings are included in the header
of the export file for result data. If disabled, only the pure result data from the selected
traces and tables is exported.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Manual operation:
See
" Include Instrument & Measurement Settings "
FORMat:DEXPort:TRACes
<Selection>
This command selects the data to be included in a data export file (see
Parameters:
<Selection>
SINGle | ALL
SINGle
Only a single trace is selected for export, namely the one speci-
fied by the
command.
ALL
Selects all active traces and result tables (e.g. Result Summary,
marker peak list etc.) in the current application for export to an
ASCII file.
The <trace> parameter for the
command is ignored.
*RST:
SINGle
Manual operation:
See
" Export all Traces and all Table Results "
MMEMory:STORe<n>:SPECtrogram
<FileName>
This command exports spectrogram data to an ASCII file.
The file contains the data for every frame in the history buffer. The data corresponding
to a particular frame begins with information about the frame number and the time that
frame was recorded.
Note that, depending on the size of the history buffer, the process of exporting the data
can take a while.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S
FSV3000/ FSVA3000 base unit user manual.
Retrieving Results