Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
273
User Manual 1178.8536.02 ─ 01
Suffix:
<n>
.
Parameters:
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR:SGR 'Spectrogram'
Copies the spectrogram data to a file.
Manual operation:
See
" Export Spectrogram to ASCII File "
MMEMory:STORe<n>:TRACe
<Trace>, <FileName>
This command exports trace data from the specified window to an ASCII file.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S
FSV3000/ FSVA3000 base unit user manual.
Suffix:
<n>
.
Parameters:
<Trace>
Number of the trace to be stored
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR1:TRAC 1,'C:\TEST.ASC'
Stores trace 1 from window 1 in the file
TEST.ASC
.
Manual operation:
See
" Export Trace to ASCII File "
8.8.4
Retrieving Marker Results
The following commands are required to retrieve the results of markers.
Useful commands for retrieving marker results described elsewhere:
●
●
●
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:X?
●
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:Y?
Retrieving Results