Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
244
User Manual 1178.8536.02 ─ 01
CALCulate<n>:MARKer<m>:SGRam:XY:MINimum[:PEAK]
CALCulate<n>:MARKer<m>:SPECtrogram:XY:MINimum[:PEAK]
This command moves a marker to the minimum level of the spectrogram.
Suffix:
<n>
.
<m>
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:ABOVe
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum:ABOVe
This command moves a marker vertically to the next lower peak level for the current
frequency.
The search includes only frames above the current marker position. It does not change
the horizontal position of the marker.
Suffix:
<n>
.
<m>
Manual operation:
See
" Search Mode for Next Peak in Y-Direction "
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:BELow
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum:BELow
This command moves a marker vertically to the next lower peak level for the current
frequency.
The search includes only frames below the current marker position. It does not change
the horizontal position of the marker.
Suffix:
<n>
.
<m>
Manual operation:
See
" Search Mode for Next Peak in Y-Direction "
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:NEXT
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum:NEXT
This command moves a marker vertically to the next lower peak level for the current
frequency.
The search includes all frames. It does not change the horizontal position of the
marker.
Suffix:
<n>
.
<m>
Manual operation:
See
" Search Mode for Next Peak in Y-Direction "
I/Q Analysis