Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
187
User Manual 1178.8536.02 ─ 01
POSitive
Triggers when the signal rises to the trigger level (rising edge).
NEGative
Triggers when the signal drops to the trigger level (falling edge).
*RST:
POSitive
Example:
TRIG:SLOP NEG
Manual operation:
See
TRIGger[:SEQuence]:SOURce
<Source>
This command selects the trigger source.
For details on trigger sources see
Note on external triggers:
If a measurement is configured to wait for an external trigger signal in a remote control
program, remote control is blocked until the trigger is received and the program can
continue. Make sure this situation is avoided in your remote control programs.
Parameters:
<Source>
IMMediate
Free Run
EXTernal
Trigger signal from the "Trigger Input" connector.
Trigger signal from the "Trigger In" connector.
If power splitter mode is active, this parameter activates the
"EXT TRIGGER INPUT" connector on the oscilloscope. Then
the R&S
FSV/A triggers when the signal fed into the "EXT TRIG-
GER INPUT" connector on the oscilloscope meets or exceeds
the specified trigger level.
EXT2
Trigger signal from the "Trigger Input/Output" connector.
Note: Connector must be configured for "Input".
Trigger signal from the "Trigger AUX" connector.
RFPower
First intermediate frequency
(Frequency and time domain measurements only.)
IFPower
Second intermediate frequency
IQPower
Magnitude of sampled I/Q data
For applications that process I/Q data, such as the I/Q Analyzer
or optional applications.
TIME
Time interval
Configuring I/Q Analyzer Measurements