Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
148
User Manual 1178.8536.02 ─ 01
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Return values:
<State>
DETected | NDETected
Example:
//Query connector state
PROB3:SET:STAT?
Usage:
Query only
[SENSe:]PROBe<pb>:SETup:TYPE?
Queries the type of the probe.
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Return values:
<Type>
String containing one of the following values:
–
"None" (no probe detected)
–
"active differential"
–
"active single-ended"
–
"active modular"
Example:
//Query probe type
PROB3:SET:TYPE?
Usage:
Query only
Manual operation:
See
8.4.1.4
External Generator Control
External generator control commands are available if the R&S
FSV/A External Genera-
tor Control option (R&S
FSV3-B10) is installed.
For each measurement channel one external generator can be configured. To switch
between different configurations define multiple measurement channels.
For more information on external generator control see
Chapter 4.3, "Basics on Exter-