Remote Commands for Bluetooth Measurements (K8)
R&S
®
FSV-K8
77
Operating Manual 1176.7610.02 ─ 02
Note:
This command is only available with active In–band Spurious Emissions measurement
(see
on page 96). With all other measure-
ments this command causes a query error.
Suffix:
<n>
.
irrelevant
Example:
For the first steps refer to
chapter 4.14, "Programming Exam-
CONF:BTO:MEAS IBS
Activates the In–band Spurious Emissions measurement.
CONF:BTO:CHAN 7
Adjust the TX channel number.
INIT;*WAI
Starts the measurement with synchronization.
CALC:BTO:IBS?
Reads the results of the In–band Spurious Emissions measure-
ment.
Usage:
Query only
Mode:
BT
Manual operation:
See
CALCulate<n>:BTOoth:IBSemissions:EXCeptions?
This command reads the number of results that exceeded the specified limits in the In–
band Spurious Emissions measurement.
Note:
This command is only available with active In–band Spurious Emissions measurement
(see
on page 96). With all other measure-
ments this command causes a query error.
Suffix:
<n>
.
irrelevant
Example:
For the first steps refer to
chapter 4.14, "Programming Exam-
CONF:BTO:MEAS IBS
Activates the In–band Spurious Emissions measurement.
CONF:BTO:CHAN 7
Adjust the TX channel number.
INIT;*WAI
Starts the measurement with synchronization.
CALC:BTO:IBS:EXC?
Reads the number of results that exceeded the specified limits.
Usage:
Query only
Mode:
BT
CALCulate subsystem