Operating Manual 1176.7561.02 ─ 06
The following table shows the required demodulation bandwidth for an error less than
0.5 dB up to a maximum AF frequency.
deemphasis
25 µs
50 µs
75 µs
750 µs
max. AF frequency
25 kHz
12 kHz
8 kHz
800 Hz
required demodulation bandwidth
≥ 200 kHz
≥ 100 kHz
≥ 50 kHz
≥ 6.4 kHz
For higher AF frequencies the demodulation bandwidth must be increased.
Remote command:
[SENSe:]FILTer<n>:DEMPhasis[:STATe]
[SENSe:]FILTer<n>:DEMPhasis:TCONstant
SFM:
[SENSe:]SFM:<ChannelType>:FILTer:DEMPhasis:STATe
[SENSe:]SFM:<ChannelType>:FILTer:DEMPhasis:TCONstant
All AF Filter Off ← AF Filter
Disables all specified AF Filters.
Remote command:
Demod Filter
By default, a flat demodulation filter is used in Analog Demodulation mode. However, in
order to optimize the settling behaviour of the filter, a Gaussian filter can be used
instead.
For details on sample rates, measurement times and trigger offsets for various demod-
ulation bandwidths when using a Gaussian filter, see
Measurement Time and Trigger Offset"
Remote command:
2.2.7
Softkeys of the Sweep Menu – SWEEP Key (Analog Demodulation)
The following table shows all softkeys available in the "Sweep" menu in "Analog
Demodulation" mode ([SWEEP] key). It is possible that your instrument configuration
does not provide all softkeys. If a softkey is only available with a special option, model
or (measurement) mode, this information is delivered in the corresponding softkey
description.
.................................................................................................56
Softkeys of the Analog Demodulation option