Ref Level
Opens an edit dialog box to enter the reference level in the current unit (dBm, dBµV,
etc).
The reference level is the maximum value the AD converter can handle without distor-
tion of the measured value. Signal levels above this value will not be measured cor-
rectly, which is indicated by the "IFOVL" status display.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
AF Range
Only available for AM/FM/PM measurements (see
For details refer to the "AF Range" softkey of the main menu (see
Range
Only available for RF measurements (see
Chapter 2.1.10, "Measurement Result Dis-
Opens a submenu to define the level display range.
Range Log 100 dB ← Range
Sets the level display range to 100 dB.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y:
Display range:
DISP:WIND:TRAC:Y 100DB
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
Range Log 50 dB ← Range
Sets the level display range to 50 dB.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y:
Display range:
DISP:WIND:TRAC:Y 50DB
, see
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]