Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
39
Operating Manual 1176.7561.02 ─ 06
Phase Wrap On/Off ← PM
Activates/deactivates the phase wrap.
On
The phase will be displayed in the range ±180° (± Π). For example, if the phase e180°,
360° is subtracted from the phase value, with the display thus showing >-180°.
Off
The phase will not be wrapped.
This softkey is available in the PM signal displays.
Remote command:
CALC:FORM PHAS
(see
RF Power
Selects RF power as the modulation type, changes the signal display, and opens a
submenu to set the measurement configuration.
In single sweep mode, the data is determined from the current I/Q data set, i.e. a
change to a different type does not trigger a new measurement.
This menu is also displayed when you press the [MEAS CONFIG] key after changing
the modulation type.
Remote command:
CALC:FEED 'XTIM:RFPower'
(see
Display Config ← RF Power
See
Select Trace ← RF Power
See
Demod BW ← RF Power
See
Meas Time ← RF Power
See
AF Filter ← RF Power
See
AF Range ← RF Power
See
Time Domain Zoom ← RF Power
See
Squelch ← RF Power
See
Squelch Level ← RF Power
See
Softkeys of the Analog Demodulation option