Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
127
Operating Manual 1176.7561.02 ─ 06
Example:
ADEM:SET 8MHz,32000,EXT,POS,-500,30
Sets up demodulator parameters
ADEM:FM AVER,MAXH,MINH
Selects FM results to be measured
ADEM:AM:REL WRIT,OFF,OFF
Selects AM results to be measured
ADEM ON
Switches on demodulator
INIT; WAI
Starts measurement and waits for sync
FORM ASC
Selects output format
ADEM:FM:RES? AVER
Reads FM average results
ADEM:FM:RES? MAXH
Reads FM max hold results
ADEM:FM:RES? MINH
Reads FM min hold results
ADEM:AM:REL:RES? WRIT
Reads current AM result data
Usage:
Query only
Mode:
ADEMOD
[SENSe:]ADEMod:AM:RELative:AFSPectrum[:TYPE]
<TraceMode>
This command selects the AF spectrum result types of the AM-demodulated signal to
be measured simultaneously.
Note:
in FM stereo mode (option K7S), only those traces can be measured that are
currently displayed in at least one screen.
Parameters:
<TraceMode>
<TraceMode1>, <TraceMode2>, <TraceMode3>, <Trace-
Mode4>, <TraceMode5>, <TraceMode6>
WRITe | AVERage | MAXHold | MINHold | VIEW | OFF
For details on trace modes see
*RST:
WRITe,OFF,OFF,OFF,OFF,OFF (FM-
Stereo:OFF,OFF,OFF,OFF,OFF,OFF)
Example:
ADEM:AM:REL:AFSP AVER,MAXH,MINH
Determines average, maximum and minimum value simultane-
ously
ADEM:AM:REL:AFSP WRIT,OFF,OFF
Determines only current measurement results
ADEM:AM:REL:AFSP OFF,OFF,OFF
Switches off calculation of the AF spectrum
Mode:
ADEMOD
Remote Commands of the Analog Demodulation (R&S
FSV–K7)