Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
108
Operating Manual 1176.7561.02 ─ 06
The corresponding marker is activated before or switched to marker mode, if neces-
sary.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
If the analog demodulator (option Analog Demodulation, R&S
FSV–K7) is activated,
the query result is output in the following units:
Result display
Output unit
AM
%
FM
Hz
PM
RF
dB (Range Log or Range Linear %)
% (Range Linear dB)
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Return values:
<Result>
The measured value of the selected marker is returned.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK2 ON
Switches marker 2.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:Y?
Outputs the measured value of marker 2.
Usage:
Query only
Manual operation:
See
"Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
2.3.3
DISPlay Subsystem (Analog Demodulation, R&S
FSV–K7)
The DISPLay subsystem controls the selection and presentation of textual and graphic
information as well as of measurement data on the display.
......................................................................................... 109
DISPlay[:WINDow<n>][:SUBWindow<1|2>]:SELect
........................................................... 109
.................................................................................... 110
DISPlay[:WINDow<n>]:TRACe<t>:MODE
.........................................................................110
DISPlay[:WINDow<n>]:TRACe<t>[:STATe]
........................................................................111
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
....................................................................111
Remote Commands of the Analog Demodulation (R&S
FSV–K7)