Remote Control
R&S
®
FSPN
425
User Manual 1179.4363.02 ─ 01
11.7.4
Marker
........................................................................................ 425
....................................................................................... 430
...............................................................................................432
.................................................................................. 436
11.7.4.1
Individual Marker Setup
CALCulate<n>:DELTamarker<m>:AOFF
...........................................................................425
CALCulate<n>:DELTamarker<m>:LINK
............................................................................ 425
CALCulate<n>:DELTamarker<ms>:LINK:TO:MARKer<md>
................................................426
CALCulate<n>:DELTamarker<m>:MODE
......................................................................... 426
CALCulate<n>:DELTamarker<m>:MREFerence
................................................................ 427
CALCulate<n>:DELTamarker<m>[:STATe]
........................................................................ 427
CALCulate<n>:DELTamarker<m>:TRACe
.........................................................................427
................................................................................. 428
..................................................................................428
CALCulate<n>:MARKer<ms>:LINK:TO:MARKer<md>
.......................................................428
CALCulate<n>:MARKer<m>[:STATe]
............................................................................... 429
................................................................................429
........................................................................................ 430
CALCulate<n>:DELTamarker<m>:AOFF
This command turns off
all
delta markers.
Suffix:
<n>
.
<m>
irrelevant
Example:
CALC:DELT:AOFF
Turns off all delta markers.
CALCulate<n>:DELTamarker<m>:LINK
<State>
This command links delta marker <m> to marker 1.
If you change the horizontal position (x-value) of marker 1, delta marker <m> changes
its horizontal position to the same value.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
Common Analysis and Display Functions