Remote Control
R&S
®
FSPN
418
User Manual 1179.4363.02 ─ 01
Prerequisites for this command
●
Turn on trace smoothing (
DISPlay[:WINDow<n>]:TRACe<t>:SMOothing[:
Suffix:
<n>
.
1..n
<t>
1..n
Parameters:
<Aperture>
<numeric value>
Range:
1 to 20
*RST:
1
Default unit: PCT
Example:
See
DISPlay[:WINDow<n>]:TRACe<t>:SMOothing[:
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:SMOothing[:STATe]
<State>
This command turns trace smoothing for a specific trace on and off.
When you turn on trace smoothing, you can define the smoothing magnitude with
DISPlay[:WINDow<n>]:TRACe<t>:SMOothing:APERture
.
Suffix:
<n>
.
1..n
<t>
1..n
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
Depends on the trace.
Example:
//Turns on trace 3 (average mode based on 10 measurements)
to display the AM noise results
DISP:TRAC3:MODE AVER
SWE:COUN 10
DISP:TRAC3:RES AM
//Smooth the trace by a magnitude of 10
%
DISP:TRAC3:SMO ON
DISP:TRAC3:SMO:APER 10
//Remove spurs from trace 3
DISP:TRAC3:SPUR:SUPP ON
DISP:TRAC3:SPUR:THR 5
Manual operation:
See
Common Analysis and Display Functions