Parameters:
<YScalUnit>
DEG | RAD
*RST:
RAD
Example:
//Select unit for phase results
DISP:TRAC:Y:UNIT DEG
Manual operation:
See
"Scaling the y-axis of the phase diagram"
11.6.9.6
Settling Time
CALCulate<n>:SETTling:TIME:STATe
.............................................................................. 398
CALCulate<n>:SETTling:TIME:DIAGram
..........................................................................398
CALCulate<n>:SETTling:TIME:TRACe
.............................................................................398
CALCulate<n>:SETTling:TIME:FREQuency:LIMit:LOWer
...................................................399
CALCulate<n>:SETTling:TIME:FREQuency:LIMit:UPPer
................................................... 399
CALCulate<n>:SETTling:TIME:RESult?
........................................................................... 399
CALCulate<n>:SETTling:TIME:STATe
<State>
This command defines the state of settling time measurement.
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC1:SETT:TIME:STAT OFF
CALCulate<n>:SETTling:TIME:DIAGram
<Diagram>
This command selects the diagram for settling time calculation.
Suffix:
<n>
.
irrelevant
Parameters:
<Diagram>
Number of the diagram
Example:
CALC:SETT:TIME:DIAG 1
CALCulate<n>:SETTling:TIME:TRACe
<Trace>
This command selects the trace for settling time calculation.
Suffix:
<n>
.
irrelevant
Parameters:
<Trace>
TRACe1 | TRACe2 | TRACe3 | TRACe4 | TRACe5 | TRACe6