Remote Control
R&S
®
FSPN
363
User Manual 1179.4363.02 ─ 01
[SENSe:]LIST:RANGe<r>:FREQuency:STOP?
This command queries the stop frequency offset of a half decade.
Suffix:
<r>
.
1..n
Return values:
<Frequency>
<numeric value>
Default unit: Hz
Example:
//Query stop frequency offset for fifth half decade
LIST:RANG5:FREQ:STOP?
would return, e.g.
300000
Usage:
Query only
Manual operation:
See
"Working with the half decade configuration table"
[SENSe:]LIST:RANGe<r>:XCOunt
<Count>
This command defines the number of cross-correlation operations done in a half dec-
ade.
The application of the command depends on the half decade configuration mode
(
):
●
Automatic mode: The number of cross-correlations can only be queried, because it
is calculated based on an RBW factor.
●
Manual mode: The number of cross-correlations can be defined directly as an
absolute value.
Suffix:
<r>
.
1..n
Parameters:
<Count>
<numeric value> (integer only)
*RST:
Automatic configuration mode is on.
Default unit: ---
Example:
//Define 50 cross-correlation operations for the fifth half decade
SWE:MODE MAN
LIST:RANG5:XCO 50
Manual operation:
See
"Defining cross-correlation parameters"
See
"Working with the half decade configuration table"