Remote Control
R&S
®
FSPN
342
User Manual 1179.4363.02 ─ 01
Example:
//Define a maximum voltage of 5 V
SOUR:VOLT:AUX:LEV:LIM:HIGH 5
Manual operation:
See
SOURce:VOLTage:AUX:LEVel:LIMit:LOW
<Voltage>
This command defines the minimum voltage that may be supplied by the V
aux
source.
Parameters:
<Voltage>
<numeric value>
Range:
-10 to 10
*RST:
0
Default unit: V
Example:
//Define a minimum voltage of -5 V
SOUR:VOLT:AUX:LEV:LIM:LOW -5
Manual operation:
See
SOURce:VOLTage:AUX:LEVel[:STATe]
<State>
This command turns the auxiliary voltage source (V
aux
) on and off.
Note that DC power is actually supplied only if you additionally activate the outputs in
general.
Prerequisites for this command
●
).
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF
Example:
//Turns on the V
aux
source.
SOUR:VOLT:AUX:LEV ON
Manual operation:
See
"Turning the output on and off"
SOURce:VOLTage:CHANnel:COUPling
<State>
This command couples or decouples the DC power configuration across measurement
channels.
Parameters:
<State>
ON | 1
DC power configuration is the same across all measurement
channels.
OFF | 0
DC power configuration is different for each measurement chan-
nel.
*RST:
ON
Common Measurement Settings