General Instrument Setup
R&S
®
FSPN
238
User Manual 1179.4363.02 ─ 01
Once the self-test is started, all modules are checked consecutively and the test result
is displayed. You can abort a running test.
In case of failure a short description of the failed test, the defective module, the associ-
ated value range and the corresponding test results are indicated.
A running Sequencer process is aborted when you start a self-test.
If you start a self-test remotely, then select the "Local" softkey while the test is still run-
ning, the instrument only returns to the manual operation state after the test is comple-
ted. In this case, the self-test cannot be aborted.
Remote command:
DIAGnostic:SERVice:STESt:RESult?
9.5.3
Calibration Signal Display
Access
: [Setup] > "Service" > "Calibration Signal"
Alternatively to the RF input signal from the front panel connector you can use the
instrument's calibration signal as the input signal, for example to perform service func-
tions on.
Service Functions