General Instrument Setup
R&S
®
FSPN
213
User Manual 1179.4363.02 ─ 01
Await Warm-Up Operation before Self Alignment
....................................................... 214
Shut down Device after Self Alignment
.......................................................................214
..........................................................................................214
Starting Touch Screen Alignment
...............................................................................214
Start Self Alignment
Starts recording correction data for the instrument. If the correction data acquisition
fails or if the correction values are deactivated, a corresponding message is displayed
in the status field.
For details, see
Chapter 9.1.1, "Basics on Alignment"
Note:
A running Sequencer operation is aborted when you start a self-alignment.
During self-alignment, do not connect a signal to the RF input connector. Running a
self-alignment with a signal connected to the RF input can lead to false measurement
results.
Remote command:
Abort Self Alignment
As long as the self-alignment data is being collected, the procedure can be canceled
using the "Abort Self Alignment" button.
Note:
If you start a self-alignment remotely, then select the "Local" softkey while the
alignment is still running, the instrument only returns to the manual operation state
after the alignment is completed. In this case, you cannot abort a self-alignment man-
ually.
Alignment