Common Measurement Settings
R&S
®
FSPN
116
User Manual 1179.4363.02 ─ 01
The table contains the following values.
●
Start Offset
Start frequency offset of the half decade.
Depends on the
●
Stop Offset
Stop frequency offset of the half decade.
Depends on the
●
RBW
Resolution bandwidth applied to the corresponding half decade during the mea-
surement.
Depends on the
Custom RBWs become available when you are using manual configuration mode.
●
XCORR
Shows the number of cross-correlation operations applied in the corresponding half
decade.
Depends on the
Custom cross-correlation operations become available when you are using manual
configuration mode.
Remote command:
Configuration mode:
Start offset:
[SENSe:]LIST:RANGe<r>:FREQuency:STARt?
Stop offset:
[SENSe:]LIST:RANGe<r>:FREQuency:STOP?
RBW:
[SENSe:]LIST:RANGe<r>:BWIDth[:RESolution]
XCORR:
6.5.4
Integrated Measurement Configuration
Access
: "Overview" > "Noise Config" > "Integrated Measurements"
You can control the integration ranges and data source (traces) that the calculations for
integrated measurement results are based on.
For integrated measurements, the raw measurement data are used instead of the data
smoothed by the
The remote commands required to configure residual calculations are described in
Chapter 11.6.6.2, "Residual Calculation Configuration"
Selecting the trace for residual calculations
................................................................ 117
Defining the integration range
..................................................................................... 117
Working with weighting filters
...................................................................................... 117
Noise Measurement Configuration