Remote Commands
R&S
®
FSMR3000
678
User Manual 1179.0116.02 ─ 02
Command
Known Differences
MKACT
Only marker 1 is supported as the active marker.
MKBW
Default value
MKPT
Step size
MKPX
Step size
OL?
Storage of instrument settings:
80 characters are returned as information on the instrument settings.
The contents of the 80 characters returned does not correspond to the original data con-
tents of the 8566A/8568A family.
OL
Readout of instrument settings:
The 80 characters read by means of OL? are accepted as information on the correspond-
ing instrument settings.
The contents of the 80 characters read does not correspond to the original data contents
of the 8566A/8568A family.
RB
Default value, range, step size
RL
Default value, step size
RLPOS
Adapts the position of the reference level even if the tracking generator normalization is
not active.
RQS
Supported bits:
1 (Units key pressed)
2 (End of Sweep)
3 (Device error)
4 (Command complete)
5 (Illegal command)
12.11.2.4
Model-Dependent Default Settings
If the GPIB language is switched over to an 85xx model, the GPIB address is automati-
cally switched over to 18 provided that the default address of the R&S
FSMR3 (20) is
still set. If a different value is set, this value is maintained. Upon return to SCPI, this
address remains unchanged.
The following table shows the default settings obtained after a change of the GPIB lan-
guage and for the commands IP, KST and RESET:
Model
# of Trace
Points
Start Freq.
Stop Freq.
Ref Level
Input Coupling
8566A/B
1001
2 GHz
22 GHz
0 dBm
AC
8568A/B
1001
0 Hz
1.5 GHz
0 dBm
AC
8560E
601
0 Hz
2.9 GHz
0 dBm
AC
8561E
601
0 Hz
6.5 GHz
0 dBm
AC
8562E
601
0 Hz
13.2 GHz
0 dBm
AC
Emulating Other Instruments' Commands