Remote Commands
R&S
®
FSMR3000
440
User Manual 1179.0116.02 ─ 02
In single measurement mode, the application stops the measurement and calculates
the average after the average count has been reached.
Example:
SWE:COUN 64
Sets the number of measurements to 64.
INIT:CONT OFF
Switches to single measurement mode.
INIT;*WAI
Starts a measurement and waits for its end.
Manual operation:
See
[SENSe:]SWEep[:WINDow<n>]:POINts
<SweepPoints>
This command defines the number of measurement points to analyze after a measure-
ment.
Suffix:
<n>
.
Example:
SWE:POIN 251
Manual operation:
See
CALCulate<n>:FEED
<Evaluation>
Selects result type for evaluation.
Suffix:
<n>
.
1..n
Parameters:
<Evaluation>
string
'XTIM:RFP:BARG'
Signal Summary
'XTIM:AC:TDOM'
Audio Timedomain
'XTIM:AC:SPEC'
Audio Spectrum
12.5.6
Triggering
The following remote commands are required to configure a triggered measurement in
a remote environment.
Configuring the Measurement