Network Operation and Remote Control
R&S
®
FSMR3000
324
User Manual 1179.0116.02 ─ 02
STATus:QUEStionable
15
14
13
12
11
10
9
8
7
6
5
4
3
0
1
2
not used
HCOPy in progress
CALibrating
15
14
13
12
11
10
9
8
7
6
5
4
3
0
1
2
not used
CALibration (=UNCAL)
POWer
LMARGin
LIMit
FREQuency
ACPLimit
15
14
13
12
11
10
9
8
7
6
5
4
3
0
1
2
ADJ UPPer FAIL
ADJ LOWer FAIL
ALT1 UPPer FAIL
ALT1 LOWer FAIL
ALT2 UPPer FAIL
ALT2 LOWer FAIL
not used
15
14
13
12
11
10
9
8
7
6
5
4
3
0
1
2
not used
15
9
8
7
6
5
4
3
0
1
2
7
6
5
4
3
0
1
2
7
6
5
4
3
0
1
2
Power On
User Request
Command Error
Execution Error
Device Dependent
Error
Query Error
Operation Complete
RQS/MSS
ESB
MAV
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
STAT:QUES:ACPLimit *)
STATus:OPERation
STAT:QUES:LMARgin<n> *)
STAT:QUES:POWer *)
not used
IF_OVerload
OVERload
ESR
ESE
Output
Buffer
Error/ Event
Queue
PPE
ISTflag
SRE
STB
SRQ
ALT3...11 LOWer/UPPer FAIL
TIME
15
14
13
12
11
10
9
8
7
6
5
4
3
0
1
2
LMARgin 1 FAIL
LMARgin 2 FAIL
LMARgin 3 FAIL
LMARgin 4 FAIL
LMARgin 5 FAIL
LMARgin 6 FAIL
LMARgin 7 FAIL
LMARgin 8 FAIL
not used
15
5
4
3
0
1
2
STAT:QUES:EXTended *)
not used
Sweep time too low
15
9
8
7
6
5
4
3
0
1
2
STAT:QUES:FREQuency *)
not used
OVEN COLD
LO UNLocked
External REFerence
STAT:QUES:LIMit<n> *)
LIMit 1 FAIL
LIMit 2 FAIL
LIMit 3 FAIL
LIMit 4 FAIL
LIMit 5 FAIL
LIMit 6 FAIL
LIMit 7 FAIL
LIMit 8 FAIL
DIQ
& logic AND
logic OR
of all bits
TEMPerature
15
9
8
7
6
5
4
3
0
1
2
not used
STAT:QUES:TEMPerature *)
Frontend temp. error
INPut_overload
Range completed
SWEeping
MEASuring
Waiting for TRIGger
*)
one register for
each channel
SYNC
Multi-Standard Capture
Finish
EXTended
15
5
4
3
0
1
2
not used
STAT:QUES:TIME *)
INFO
15
5
4
3
0
1
2
STAT:QUES:EXTended:INFO *)
not used
INFO
MESSage
WARNing
ERRor
FATal
...
...
...
...
...
...
CACLR FAIL
GAP ACLR FAIL
PCALibration
15
14
13
12
11
10
9
8
7
6
5
4
3
0
1
2
UNCAL
CALibrated
RECAL needed
CAL ABS POWER active
Power Sensor Overload
No Sensor
not used
STATus:OPERation:PCALibration
RF Level out of Tolerance
Figure 11-1: Graphical overview of the R&S
FSMR3 status registers hierarchy
All status registers have the same internal structure.
11.2.2
Contents of the Status Registers
In the following sections, the contents of the status registers are described in more
detail.
STATus:QUEStionable:SYNC register
The
STATus:QUEStionable:SYNC
register is used by the R&S
FSMR3 applications
and is described in the individual sections (manuals) for each application.
Status Reporting System