Data Management
R&S
®
FSMR3000
252
User Manual 1179.0116.02 ─ 02
These functions are only available if no measurement is running.
In particular, if a continuous measurement is running, the import/export functions are
not available.
........................................................................... 252
........................................................................................252
Export
Access
: "Save/Recall" > Export
Opens a submenu to configure data export.
Trace Export Configuration ← Export
Opens the "Traces" dialog box to configure the trace and data export settings.
I/Q Export ← Export
Opens a file selection dialog box to define an export file name to which the I/Q data is
stored. This function is only available in single sweep mode.
It is only available in applications that process I/Q data, such as the I/Q Analyzer or
other optional applications.
For details, see the description in the R&S
FSMR3 I/Q Analyzer User Manual ("Import-
ing and Exporting I/Q Data").
Note:
Storing large amounts of I/Q data (several Gigabytes) can exceed the available
(internal) storage space on the R&S
FSMR3. In this case, it can be necessary to use
an external storage medium.
Note:
Secure user mode.
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
File Explorer ← I/Q Export ← Export
Opens the Microsoft Windows File Explorer.
Remote command:
not supported
9.5
Creating Screenshots of Current Measurement
Results and Settings
To document the graphical results and the most important settings for the currently per-
formed measurement, you can create a screenshot of the current display. Screenshots
can either be printed or stored to a file.