background image

R&S

©

FPC-Z10 Teaching Kit 

Getting Started 1178.8436.02 – 0

4

Attenuator 

Both the upconverter and downconverter have an attenuator. Both signal paths have a DIP 
switch to select the attenuation (7). 

Move one of the switches to its "Off" position (the side of the inscription). 
Each switch corresponds to a specific attenuation. The switches are 
additive.  
Example: If you move the "8 dB" and the "2 dB" switches to their "Off" position you get an 
attenuation of 10 dB. Thus, you can define an attenuation between 0 and 31.5 dB.

 

Switches 7 and 8 have no function. 

Local oscillator configuration 

The local oscillator (LO) can supply two different frequencies. The LO frequency is used by the 
mixer to change the frequency to the intermediate frequency (IF). You can select the frequency 
with the jumper (1). 

Position 1:  LO frequency of 233.5 MHz. 
Position 2:  LO frequency of 636.5 MHz. 

Select this frequency if you are using the internal bandpass 

 

 

filter.

 

Synthesizer configuration 

The synthesizer provides the frequency for the local oscillator. You can select the 
synthesizer state with jumper (

2

Position 1:  Turns on the synthesizer. 
Position 2:  Resets the synthesizer. 

A synthesizer reset is necessary when you change the LO  
frequency. After a reset, you have to switch the jumper to  
position 1 again.

 

Calibration kit 

You can calibrate VNA measurements with the calibration kit. It provides all necessary 
calibration standards: a short, an open, a 50 Ohm match and a through. 

The setting of jumper (6) selects the usage of connectors (C), (D) and (E) (labeled "Short" and 
"Open"). 

Position 1:  Selects a short and open calibration. 
Position 2:  Selects a through calibration.

 

For a comprehensive description of the calibration procedures, refer to the R&S

®

FPC user 

manual. 

Summary of Contents for FPC-Z10 Teaching Kit

Page 1: ...R S FPC Z10 Teaching Kit Getting Started 1178843602 Version 04 Getting Started...

Page 2: ...164 Email info rohde schwarz com Internet www rohde schwarz com Subject to change Data without tolerance limits is not binding R S is a registered trademark of Rohde Schwarz GmbH Co KG Trade names are...

Page 3: ...ies to IEC UL EN 60950 1 or IEC UL EN 62368 1 Make sure that the RF signals you apply have a maximum level of 10 dBm and a voltage of 0 V DC Connected equipment must also comply to IEC UL EN 60950 1 o...

Page 4: ...al routing 4 Jumper for IF PLL power supply selection 5 Power supply selection 6 Calibration standard selection 7 Attenuator for up downconverter 8 Baseband state Connectors A Power supply micro USB B...

Page 5: ...itry the rating is 5 V DC at 450 mA Use the USB cable delivered with the teaching kit to connect it to another USB interface You can use the USB interfaces on the R S FPC as the power source for examp...

Page 6: ...apply a signal make sure not to overload the inputs and outputs max 0 dBm You can bypass or close any of the inputs and outputs by setting the corresponding jumpers Here s an example for the downconve...

Page 7: ...ency with the jumper 1 Position 1 LO frequency of 233 5 MHz Position 2 LO frequency of 636 5 MHz Select this frequency if you are using the internal bandpass filter Synthesizer configuration The synth...

Page 8: ...upply You can measure the noise of the PLL power supply by measuring the signal on the X105 output The switching converter produces more RF noise than the linear converter Service and maintenance The...

Reviews: