Configuration
R&S
®
ESW-K58
33
User Manual 1179.0880.02 ─ 03
5.5.2
Trigger input and output settings
Access
: "Overview" > "Trigger" > "Trigger In/Out"
Trigger 2/3
Defines the usage of the variable Trigger Input/Output connectors, where:
"Trigger 2": Trigger Input/Output connector on the front panel
"Trigger 3": Trigger 3 Input/Output connector on the rear panel
(Trigger 1 is INPUT only.)
Providing trigger signals as output is described in detail in the R&S
ESW User Manual.
"Input"
The signal at the connector is used as an external trigger source by
the R&S
ESW. Trigger input parameters are available in the "Trigger"
dialog box.
"Output"
The R&S
ESW sends a trigger signal to the output connector to be
used by connected devices.
Further trigger parameters are available for the connector.
Remote command:
OUTPut<up>:TRIGger<tp>:DIRection
5.6
Measurement settings
Access
: "Meas Config" > "Meas Settings"
Measurement settings