Remote Control
R&S
®
ESR
993
User Manual 1175.7068.02 ─ 12
For FSP (=FSV), the standard parameter coupling of the instrument is used. As a
result, in most cases a shorter sweeptime is used than in case of HP.
This command is only available if a HP language is selected using
.
Parameters:
<CouplingType>
HP | FSP
*RST:
FSP
Example:
SYSTem:HPC HP
Usage:
SCPI confirmed
Manual operation:
See
SYSTem:IFGain:MODE
<Mode>
Configures the internal IF gain settings in HP emulation mode due to the application
needs. This setting is only taken into account for resolution bandwidth < 300 kHz and is
only available if a HP language is selected using
.
Parameters:
<Mode>
NORM | PULS
NORM
Optimized for high dynamic range, overload limit is close to ref-
erence level.
PULS
Optimized for pulsed signals, overload limit up to 10 dB above
reference level.
*RST:
NORM
Example:
SYST:IFG:MODE PULS
Usage:
SCPI confirmed
Manual operation:
See
SYSTem:TWALking
<Mode>
Defines how remote control programs are evaluated, in particular multiple commands
in one command line.
Note that the default setting is
NORM
; however, the setting defined by this command is
retained after a [preset] (*RST) operation.
Parameters:
<Mode>
NORM
SCPI-conform behaviour
The first command is always referenced to the root node. Sub-
sequent commands, however, are referenced to the same tree
level as the previous command in a message unit.