Remote Control
R&S
®
ESR
935
User Manual 1175.7068.02 ─ 12
11.9.6
Calibrating the R&S
ESR
DIAGnostic<n>:SERVice:INPut:PULSed:CFRequency
.......................................................936
DIAGnostic<n>:SERVice:INPut:PULSed:MCFRequency
.................................................... 937
DIAGnostic<n>:SERVice:INPut[:SELect]
.......................................................................... 937
DIAGnostic<n>:SERVice:STESt:RESult?
......................................................................... 937
DIAGnostic<n>:SERVice:TEMPerature:FRONt?
................................................................938
CALibration:ABORt
This command aborts the current calibration routine.
Example:
CAL:ABORt
Aborts calibration.
Usage:
Event
SCPI confirmed
CALibration[:ALL]?
This command initiates a calibration routine and queries if calibration was successful.
During the acquisition of correction data the instrument does not accept any remote
control commands.
In order to recognize when the acquisition of correction data is completed, the MAV bit
in the status byte can be used. If the associated bit is set in the Service Request
Enable Register, the instrument generates a service request after the acquisition of
correction data has been completed.
Return values:
<CalibrationState>
0
The command returns a '0' if calibration was successful.
Example:
*CLS
Resets the status management.
*SRE 16
Enables MAV bit in the Service Request Enable Register.
*CAL?
Starts the correction data recording, and then a service request
is generated.
Usage:
Query only
SCPI confirmed
CALibration:RESult?
This command returns the results collected during calibration.
System Configuration