Parameters:
<Type>
LEVel | EDGE
*RST:
EDGE
Example:
TRAC:IQ:EGAT:TYPE LEV
TRACe<n>:IQ:EVAL
<State>
This command enables or disables the IQ data evaluation mode and is a prerequisite
for the IQ Analyzer functions.
Before this command can be executed, IQ data acquistion must be enabled (see
on page 898). This command automatically disables the
IQ data evaluation mode to provide optimum performance for IQ-data acquisition.
If you start the IQ Analyzer manually by selecting the "IQ Analyzer" mode in the
R&S
command, the IQ data evaluation
mode is automatically activated.
The maximum number of samples for RF input is 200 MS. For details see
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF
ON
all functions of the IQ Analyzer are available
OFF
no traces or marker operations are available; IQ data can only
be captured, but with a better performance than in active IQ data
evaluation mode
*RST:
OFF
Example:
TRAC:IQ ON
Enables IQ data acquisition
TRAC:IQ:EVAL ON
Enables the IQ data evaluation mode.
Manual operation:
See
TRACe<n>:IQ[:STATe]
<State>
This command switches the I/Q data acquisition on or off.
I/Q data acquisition is not compatible with other measurement functions. Therefore, all
other measurement functions are switched off as soon as the I/Q measurement func-
tion is switched on. The trace display is also switched off by this command. Therefore,
all traces are set to "BLANK".
Remote Commands in I/Q Analyzer Mode