Remote Control
R&S
®
ESR
703
User Manual 1175.7068.02 ─ 12
Manual operation:
See
"Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
CALCulate<n>:DELTamarker<m>:Y
This command queries the measured value of a delta marker. The corresponding delta
marker is activated, if necessary. The output is always a relative value referred to
marker 1 or to the reference position (reference fixed active).
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
The unit of results depends on the result display and the unit you have selected.
Table 11-20: Base unit
Parameter, measuring function or result display
Output unit
DBM | DBPW | DBUV | DBMV | DBUA
dB (lin/log)
WATT | VOLT | AMPere
dB (lin), % (log)
statistics function (APD or CCDF) on
dimensionless output
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Example:
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
CALC:DELT2 ON
Switches on delta marker 2.
CALC:DELT2:Y?
Outputs measurement value of delta marker 2.
Manual operation:
See
"Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
11.4.8
Limit Lines
...............................................................................................703
......................................................................................... 704
11.4.8.1
Using Display Lines
................................................................................................704
..................................................................................... 704
Remote Commands in Receiver Mode