Remote Control
R&S
®
ESR
687
User Manual 1175.7068.02 ─ 12
FOURphase
R&S ESH2-Z5: four phases and protective earth are controlla-
ble.
OFF
Turns off remote control of the LISN.
TWOPhase
R&S ESH3-Z5: two phases and protective earth are controllable.
*RST:
OFF
Example:
INP:LISN TWOP
Turns automatic control of the network on and selects the R&S
ESH3-Z5 network.
Manual operation:
See
See
11.4.7
Working with Markers
Configuring Marker Functionality
.......................................................................... 687
..............................................................................................697
11.4.7.1
Configuring Marker Functionality
CALCulate<n>:MARKer<m>:COUPled[:STATe]
.................................................................687
CALCulate<n>:MARKer<m>:FUNCtion:CENTer
................................................................ 688
CALCulate<n>:MARKer<m>:FUNCtion:CSTep
..................................................................688
CALCulate<n>:MARKer<m>:FUNCtion:ZOOM
..................................................................688
CALCulate<n>:MARKer<m>:SCOupled[:STATe]
................................................................689
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
....................................................................689
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHT
..................................................................689
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
.................................................................690
DISPlay[:WINDow<n>]:MINFo:STATe
...............................................................................690
CALCulate<n>:MARKer<m>:COUPled[:STATe]
<State>
This command couples or decouples the receiver frequency to the current marker fre-
quency.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:MARK:COUP ON
Couples the frequency to the marker frequency.
Manual operation:
See
Remote Commands in Receiver Mode