Remote Control
R&S
®
ESR
670
User Manual 1175.7068.02 ─ 12
INPut:IMPedance
<Impedance>
This command selects the nominal input impedance.
75
Ω
should be selected if the 50
Ω
input impedance is transformed to a higher impe-
dance using a 75
Ω
adapter of the RAZ type (= 25
Ω
in series to the input impedance
of the instrument). The correction value in this case is 1.76 dB = 10 log (75
Ω
/50
Ω
).
Parameters:
<Impedance>
50 | 75
*RST:
50 Ω
Example:
INP:IMP 75
Manual operation:
See
11.4.4
Selecting the Bandwidth
[SENSe:]BANDwidth|BWIDth[:RESolution]
....................................................................... 670
[SENSe:]BANDwidth|BWIDth[:RESolution]:AUTO
..............................................................671
[SENSe:]BANDwidth|BWIDth[:RESolution]:TYPE
.............................................................. 671
[SENSe:]BANDwidth:IF
<Bandwidth>
This command defines the resolution bandwidth for IF analysis.
Parameters:
<Bandwidth>
Resolution filter bandwidth with a 6 dB bandwidth in the range
from 10 Hz to 100 kHz in steps of 1-3-10-....
The available bandwidths depend on the current frequency
span.
*RST:
3 kHz
Example:
BAND:IF 10KHZ
Defines a resolution bandwidth of 10 kHz.
Manual operation:
See
[SENSe:]BANDwidth|BWIDth[:RESolution]
<Bandwidth>
This command defines the resolution bandwidth.
The available bandwidths depend on the selected filter type. For more information see
Chapter 5.2.1, "Measurement Bandwidth"
A change of the resolution bandwidth automatically turns the coupling to the span off.
Parameters:
<Bandwidth>
refer to data sheet
*RST:
(AUTO is set to ON)
Remote Commands in Receiver Mode