Remote Control
R&S
®
ESR
647
User Manual 1175.7068.02 ─ 12
PHOLD
See
SCAN
See
"Querying scan characteristics"
SGRam
See
SINGLE
See
STATUS
See
"Querying scan characteristics"
Return values:
<TraceData>
For more information see tables below.
Example:
TRAC? TRACE1
Queries the level for each trace point of trace 1.
TRACe<n>:IF[:DATA]
<ResultType>
This command queries current trace data and measurement results for IF analysis.
The data format depends on
FORMat[:DATA]
.
Suffix:
<n>
.
irrelevant
Parameters:
<ResultType>
TRACE1 | ... | TRACE3
Returns the level values displayed on the corresponding trace.
Example:
TRAC:IF? TRACE1
Queries the level for each trace point of trace 1.
11.4.1.6
Fixed Frequency Scans
Commands useful to configure the fixed frequency scans described elsewhere:
●
............................................................................................... 647
[SENSe:]SCAN:TDOMain
<Time>
This command defines the measurement time for fixed frequency scans.
Parameters:
<Time>
Measurement time in seconds.
The range indicated eblow is the maximum range. The actual
range depends on the measurement time defined with
Range:
10 ms to 10000 s
Remote Commands in Receiver Mode