Spectrum Measurements
R&S
®
ESR
397
User Manual 1175.7068.02 ─ 12
Preset All Traces
Trace 2-6:
Select Max | Avg | Min
Trace 4-6:
Select Max | ClrWrite | Min
Trace 4-6:
6.3.1.3
Specifying the Trace Settings
1. To configure one or more traces, see
Chapter 6.3.1.2, "Configuring Traces"
2. To select the trace mode for the selected trace, press the softkey for the corre-
sponding trace (for details see
3. To select a detector, press the
softkey for automatic detector selection,
or press the
Chapter 6.3.1.5, "Detector Overview"
4. To change the sweep count setting, which also determines trace averaging, press
the
5. To deactivate the reset of the traces in "Min Hold" and "Max Hold" mode after some
specific parameter changes, press the
6. To copy a trace into another trace memory, press the
Upon copying, the contents of the selected memory are overwritten and the new
contents are displayed in the View mode.
7. To export the active trace in ASCII format:
a) Press the "More" softkey.
b) If necessary, press the
softkey to change the decimal separator
with floating-point numerals.
c) Press the
softkey to enter the ASCII file export name.
The active trace is saved in ASCII format on the harddisk on or an external
storage device.
6.3.1.4
Description of the Averaging Method
Averaging is carried out over the measurement points derived from the measurement
samples. Several measured values may be combined in a measurement point. This
Analysis