Spectrum Measurements
R&S
®
ESR
391
User Manual 1175.7068.02 ─ 12
Remote command:
DISP:TRAC OFF
, see
DISPlay[:WINDow<n>]:TRACe<t>[:STATe]
Hold/Cont ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Switches the reset of the traces in Min Hold, Max Hold and Average mode after some
specific parameter changes have been made on and off. The default setting is off.
Normally, the measurement is started anew after parameter changes, before the mea-
surement results are evaluated (e.g. using a marker). In all cases that require a new
measurement after parameter changes, the trace is reset automatically to avoid false
results (e.g. with span changes). For applications that require no reset after parameter
changes, the automatic reset can be switched off.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:MODE:HCONtinuous
Detector ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Opens a submenu to select the detector manually, or activate automatic selection.
Note:
When measuring spurious emissions, using this softkey automatically opens the
If a detector was selected manually, the "MAN" indicator is highlighted.
If "AUTO" is selected, the detector is defined automatically, depending on the selected
trace mode:
Trace mode
Detector
Clear Write
Auto Peak
Max Hold
Positive Peak
Min Hold
Negative Peak
Average
Sample Peak
View
–
Blank
–
Auto Select ← Detector ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Selects the best detector for the selected trace and filter mode. This is the default set-
ting.
Trace mode
Detector
Clear/Write
Auto Peak
Average
Sample
Max Hold
Max Peak
Min Hold
Min Peak
Remote command:
[SENSe:][WINDow:]DETector<trace>[:FUNCtion]:AUTO
Analysis