Registration
This section defines settings of the registration procedure.
Timer Based Registration is available with the basic signaling application R&S CMW-
KS800. All other registration types require the advanced signaling option R&S CMW-
KS810.
Fig. 4-54: Registration settings
............................................................................................561
....................................................................................... 562
Autonomous Registration (Home / Foreign SID / Foreign NID)
.................................. 562
.................................................................................................562
............................................................................................ 562
.................................................................................. 563
Timer-based Registration
Timer-based registration causes the mobile station to register at regular intervals. Its
use also allows the system to automatically deregister mobile stations that did not per-
form a successful power-down registration.
This is an autonomous registration; see
Autonomous Registration (Home / Foreign
The interval is controlled by the base station via the REG_PRD field of the System
Parameters Message or ANSI-41 System Parameters Message. If timer-based regis-
tration shall be performed, this field shall be set to a value in the range 29 to 85 inclu-
sive and the timer value is determined by the following formula:
[2
REG_PRD/4
]x0.08 s
This results in intervals from approximately 12 seconds to 3 days
Remote command: