1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
354
User Manual 1173.9511.02 ─ 11
CONFigure:EVDO:MEAS<i>:MEValuation:LIMit:ACP:LOWer:ABSolute
<ACPSetting>(10)
Defines lower limits 9 to 0 for the ACP measurement in dBm of the selected carrier
(see
ual negative offset frequencies (set via
).
The limit index 9 to 0 corresponds to the index used in manual control.
Parameters:
<ACPSetting>
Range:
-80 dBm to 10 dBm
*RST:
For offset frequencies 9 to 0 all -54 dBm, OFF
Default unit: dBm
Additional parameters: OFF | ON (disables | enables the limit
check)
Example:
See
V3.2.70
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:LIMit:ACP:UPPer:ABSolute
<ACPSetting>(10)
Defines upper limits 0 to 9 for the extended ACP measurement in dBm of the selected
carrier (see
) at the
individual positive offset frequencies (set via
).
The limit index 0 to 9 corresponds to the index used in manual control.
Parameters:
<ACPSetting>
Range:
-80 dBm to 10 dBm
*RST:
For offset frequencies 0 to 9 all -54 dBm, OFF
Default unit: dBm
Additional parameters: ON | OFF (enables | disables the limit
check)
Example:
See
V3.2.70
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:LIMit:ACP:LOWer[:RELative]
<Limit_9>, ..., <Limit_0>
Defines limits for the relative ACP in dBc of the selected carrier (see
) at the individual negative offset
).
Command Reference