1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
342
User Manual 1173.9511.02 ─ 11
CONFigure:EVDO:MEAS<i>:MEValuation:MOEXception
<MeasOnException>
Specifies whether measurement results that the R&S
CMW identifies as faulty or inac-
curate are rejected.
Parameters:
<MeasOnException> OFF | ON
ON:
Results are never rejected
OFF:
Faulty results are rejected
*RST:
OFF
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
V1.0.3.3
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:ILCMask
<LCMaskI>
Specifies the long code mask for the I branch.
Parameters:
<LCMaskI>
Range:
#H0 to #H3FFFFFFFFFF
*RST:
#H0
Example:
See
Specifying General Measurement Settings
Firmware/Software:
V1.0.3.3
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:QLCMask
<LCMaskQ>
Specifies the long code mask for the Q branch.
Parameters:
<LCMaskQ>
Range:
#H0 to #H3FFFFFFFFFF
*RST:
#H0
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
V1.0.3.3
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:RPMode
<RefPowerMode>
Sets the reference power relative to which the power (in dB) of the reverse link physi-
cal channels of both the I and Q signal are measured.
Command Reference