Remote control commands
R&S
®
AREG800A
353
User Manual 1179.3615.02 ─ 01
:SENSe<ch>[:POWer]:DISPlay:PERManent:STATe
<State>
Activates the permanent display of the measured power level results. The instrument
also indicates the sensor type, the connection, the measurement source and the offset
if set.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
:SENS1:POW:DISP:PERM:STAT ON
Turns on the permanent view.
Manual operation:
See
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?
Queries the current filter length in filter mode
AUTO
)
Return values:
<Auto>
float
Range:
1 to 65536
Example:
:SENS1:FILT:TYPE AUTO
Selects auto filter.
:SENS1:FILT:LENG:AUTO?
Queries the automatically set filter length.
Response: 1024
Usage:
Query only
Manual operation:
See
:SENSe<ch>[:POWer]:FILTer:LENGth[:USER]
<User>
Selects the filter length for
. As the filter length works as
a multiplier for the time window, a constant filter length results in a constant measure-
ment time (see also
"About the measuring principle, averaging filter, filter length, and
The R&S NRP power sensors provide different resolutions for setting the filter length,
depending on the used sensor type:
●
Resolution = 1 for R&S NRPxx power sensors
●
Resolution = 2
n
for sensors of the R&S NRP-Zxx family, with n = 1 to 16
Parameters:
<User>
float
Range:
1 to 65536
*RST:
1
SENSe, READ, INITiate and SLISt subsystems