R&S AFQ100A
Status Reporting System
1401.3084.32 5.31
E-3
Reset Values of the Status Reporting Systems
Commands
*RST
,
*DCL
and
SYSTem:PRESet
and switching on the supply voltage also affect the status
reporting system. None of the commands, except for
*RST
and
SYSTem:PRESet
influences the
functional instrument settings. In particular, DCL does not change the instrument settings.
Table 5-4
Resetting the Status Reporting System
Event
Switching on
supply voltage
DCL,SDC
Power-On-Status-
Clear
(Device Clear,
Selected Device
Clear)
*RST or
SYSTem:PRESet
STATus:PRESet *CLS
Effect 0
1
Clear
STB,ESR
– yes – – – yes
Clear
SRE,ESE
– yes – – – –
Clear
PPE
– yes – – – –
Clear
error-queue yes yes – – – yes
Clear
output
buffer yes yes yes
1) 1) 1)
Clear command processing
and input buffer
yes yes yes
– – –
Table 5-5
Resetting the Status Reporting System
1) Every command being the first in a command line, i.e. immediately following a <PROGRAM MESSAGE TERMINATOR>
clears the output buffer