With folded-out feet, do not place anything on top or underneath the instrument.
3.
WARNING!
A stack of products can fall over and cause injury. Never stack more
than three products on top of each other. Instead, mount them in a rack.
Stack as follows:
● It is best if all instruments have the same dimensions (width and length).
● The overall load on the lowest instrument must not exceed 500
N.
● With smaller instruments on top of the lowest instrument, the overall load on
the lowest instrument must not exceed 250
N.
Too many instruments
max. 500 N
Correct
Different dimensions
max. 250 N
4.
NOTICE!
Overheating can damage the product.
Prevent overheating as follows:
● Keep a minimum distance of 10
cm between the fan openings of the R&S NRX
and any object in the vicinity.
● Do not place the R&S NRX next to heat-generating equipment such as radia-
tors or other instruments.
3.1.5
Mounting in a Rack
1. Order one of the rack adapter kits designed for the R&S NRX. For the order num-
ber, see data sheet.
2. Follow the installation instructions provided with the adapter kit.
3.
NOTICE!
Insufficient airflow can cause overheating and damage the product.
Design and implement an efficient ventilation concept for the rack.
3.1.6
Connecting to the AC Power Supply
The R&S NRX can be used with different AC power voltages and adapts itself auto-
matically to them. Adjusting the R&S NRX to a particular AC supply voltage is therefore
not required. Refer to the data sheet for the requirements of voltage and frequency.
The power switch can be set to two positions: