Command reference
R&S
®
ZNA
948
User Manual 1178.6462.02 ─ 20
CALCulate:FMODel:ISD<Ph_pt>:COUPon[:STATe]
<Boolean>
For a fixture modeling with the ISD tool, this command allows to specify the ports to
which the test coupon is connected.
This has to be defined before measuring the test coupon (using
or
CALCulate:FMODel:ISD<Ph_pt>:COUPon:
CALCulate:FMODel:ISD<Ph_pt>:COUPon:MEASure:SHORt
).
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean>
1 (ON, true) if the test coupon is connected to port <Ph_pt>, 0
(OFF, false) otherwise
Example:
See
Chapter 8.2.8, "Fixture modeling"
Manual operation:
See
CALCulate:FMODel:ISD<Ph_pt>:DCEXtrapolat
<Boolean>
Defines whether the de-/embedding file generated by the ISD tool contains (extrapola-
ted) DC data.
Can only be set to ON, if a recent version of the ISD tool (from 2019-12 or later) is
used. ON correspons to
add_dc=1
in the tool's batch mode.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Boolean>
OFF (0)
No DC data added, i.e. the touchstone data reach down to the
start frequency of the instrument sweep.
ON (1)
The ISD tool adds (extrapolated) S-parameter data for f = 0.
*RST:
OFF (0)
Manual operation:
See
CALCulate:FMODel:ISD<Ph_pt>:DUT:MEASure
For a fixture modeling with the ISD tool, this command allows to measure DUT + Fix-
ture at the active ports (see
CALCulate:FMODel:ISD<Ph_pt>:DUT[:STATe]
)
Suffix:
<Ph_pt>
.
This suffix is ignored.
Example:
See
Chapter 8.2.8, "Fixture modeling"
Usage:
Event
Manual operation:
See
SCPI command reference