Command reference
R&S
®
ZNA
937
User Manual 1178.6462.02 ─ 20
Suffix:
<Ph_pt>
.
This suffix is ignored
Parameters:
<SweepControl>
DEFault | CURRent
DEFault
Use the default values of the Delta-L algorithm:
10
MHz frequency step size, 1
kHz IF bandwidth
SweepControl
Use the values of the active channel
Options:
R&S
ZNA-K231
Manual operation:
See
CALCulate:FMODel:DELT<Ph_pt>:SWEep:CONTrol:BWIDth[:RESolution]:
DREDuction
<Boolean>
Defines whether dynamic bandwidth at low frequencies is used for Delta-L measure-
ments (see
[SENSe<Ch>:]BANDwidth[:RESolution]:DREDuction
Suffix:
<Ph_pt>
.
This suffix is ignored
Parameters:
<Boolean>
ON (1)
Dynamic bandwidth at low frequencies is active (default)
OFF (0)
Dynamic bandwidth at low frequencies is inactive
Options:
R&S
ZNA-K231
Manual operation:
See
CALCulate:FMODel:DELT<Ph_pt>:SWEep:CONTrol:IFBW?
For the Delta-L measurement, the firmware can either use a default IF bandwidth, or
the active channel setting (see
CALCulate:FMODel:DELT<Ph_pt>:SWEep:
).
This query returns the resulting IF bandwidth.
Suffix:
<Ph_pt>
.
This suffix is ignored
Usage:
Query only
Options:
R&S
ZNA-K231
Manual operation:
See