GUI reference
R&S
®
ZNA
635
User Manual 1178.6462.02 ─ 20
Remote command:
Use Primed Waves
If a (two-tone)
Embedded LO mixer group delay measurements
is active, in addition to
a and b the analyzer can provide "primed" wave quantities a', b'. The unprimed wave
quantities correspond to the lower tone, the primed wave quantities correspond to the
upper tone.
Remote command:
[SENSe<Ch>:]FREQuency:OFFSet:PWAVes:STATe
Track LO
If a frequency-converting measurement is active, the analyzer can track the difference
between the expected and the actual LO frequency.
Note:
The "Track LO" functionality is now implemented within the original channel.
Before V1.80 of the analyzer firmware, each time "Track LO" was switched on, a new
(independent) channel was created. In particular, it was up to the user to ensure the
correct position in the channel sequence and the consistency of the values with the
"parent" channel.
Remote command:
SOURce<Ch>:LOTRack[:STATe]
SOURce<Ch>:LOTRack:DFRequency?
5.13.1.2
Wideband IF Gain Manual Configuration dialog
The "Wideband IF Gain Manual Configuration" dialog allows you to configure the wide-
band / narrowband IF gain settings for individual sweep segments, drive ports and
receivers. This button is enabled in "Manual"
only.
Access:
Channel – [Mode] > "Mode" > "IF Gain Mode" – "Manual Config..."
Mode softtool