Command reference
R&S
®
ZNA
1541
User Manual 1178.6462.02 ─ 20
[SENSe<Ch>:]FREQuency:OFFSet:PWAVes:STATe
<Boolean>
Enables/disables the calculation of primed wave quantities for two-tone group delay
measurements.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
ON | OFF | 0 | 1
*RST:
1
Options:
R&S
ZNA-K9
Manual operation:
See
[SENSe<Ch>:]FREQuency<Pt>:OFFSet:PWAVes
<FreqOffset>
Defines the constant receiver frequency offset for primed wave quantities a' and b' rela-
tive to the common receiver frequency.
Suffix:
<Ch>
.
Channel number
<Pt>
Physical port number
Parameters:
<FreqOffset>
Frequency offset.
Range:
0 Hz to 25 MHz
Increment: 1 Hz
*RST:
0 Hz
Default unit: Hz
Example:
FREQ:STAR 2GHz; STOP 3 GHz
Define a sweep range between 2 GHz and 3 GHz.
FREQ:OFFS:PWAV 10kHz; WAV 20kHz
Modify the receiver frequency settings.
[SENSe<Ch>:]FREQuency<Pt>:OFFSet:WAVes
<FreqOffset>
Defines the constant receiver frequency offset for unprimed wave quantities a, b rela-
tive to the common receiver frequency.
Suffix:
<Ch>
.
Channel number.
<Pt>
Test port number of the analyzer.
Parameters:
<FreqOffset>
Frequency offset.
Range:
0 Hz to 25 MHz
Increment: 1 Hz
*RST:
0 Hz
Default unit: Hz
SCPI command reference